Guidance And Manufacturer's Declaration - Electromagnetic Immunity - GE MAC 5000 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5000:
Table of Contents

Advertisement

Appendix C – Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
Immunity Test
Electrostatic discharge
(ESD)
± 6 kV contact
EN 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
±1 kV for input/output lines
EN 61000-4-4
Surge
± 1 kV differential mode
EN 61000-4-5
± 2 kV common mode
Voltage dips, short
<5% Ut (>95% dip in U
interruptions and
for 0.5 cycles
voltage variations on
40% Ut ( 60% dip in U
power supply input
for 5 cycles
lines
70% Ut ( 30% dip in U
EN 61000-4-11
for 25 cycles
<5% Ut (>95% dip in U
for 5 sec
Power frequency
(50/60 Hz) magnetic
3 A/m
field
EN 61000-4-8
C-4
The MAC 5000 is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to ensure
that the MAC 5000 is used in such an environment.
EN 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
±1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
)
<5% Ut (>95% dip in U
t
for 0.5 cycles
)
40% Ut ( 60% dip in U
t
for 5 cycles
)
70% Ut ( 30% dip in U
t
for 25 cycles
)
<5% Ut (>95% dip in U
t
for 5 sec
3 A/m
NOTE
Ut is the AC mains voltage prior to application of the test level.
MAC™ 5000 resting ECG analysis system
Compliance Level
)
t
)
t
)
t
)
t
2024917-010
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power should be that of a typical
commercial or hospital environment.
Mains power should be that of a typical
commercial or hospital environment.
Mains power should be that of a typical
commercial or hospital environment. If the
user of the MAC 5000 requires continued
operation during power mains interruptions,
it is recommended that the MAC 5000 be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should be
at levels characteristics of a typical location
in a typical commercial or hospital
environment.
Revision B