12.1.2

Running a test sequence

Starting a sequence
1. Select the desired sequence from the Immittance module control panel.
Start
2. Click
Warning
Already measured results are kept.
In reflex testing the data collection will be paused before each stimulus above 100 dB HL, and you will
be prompted to continue.
Stopping a sequence
If a leak is detected or the probe is removed from the patient's ear, the sequence stops automatically.
1. Click
Stop
2. If needed, perform supplementary measurements. You can supplement an interrupted sequence by
deselecting the successful part and then restart the sequence testing.
12.2

Quick Check testing

You can perform Tympanometry and Reflex Screening as a Quick Check test from the
Note
The safety intensity level cannot be exceeded in screening mode.
Preparing for testing
1. Do as described in
2. Launch the Immittance module with the
3. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
4. If needed, toggle the
5. Load a test setting configured for Quick Check testing of patients in the relevant age group.
Procedure
1. To prepare the probe for Quick Check testing, press the
it into the ear canal of the patient.
Otometrics - MADSEN Zodiac
to start the selected sequence.
If the patient is troubled by the test, stop the test. The test is interrupted immediately.
to stop the sequence manually.
Preparing for testing ► 36
Ear Selector
to select the ear on which you wish to start the test.
12 Testing with Zodiac Diagnostic or Clinical - PC-based
.
Tymp + Reflex Screening
screen selected.
Start
button on the probe or shoulder pad before you insert
Tymp + Reflex Screening
screen.
87