D. Tympanometry results
The results table shows the results related to the currently selected curve:
Probe
(the probe tone in Hz)
TPP
(Tympanometric Peak Pressure)
ECV
(Equivalent Ear Canal Volume)
SA
(Static Peak Admittance), or
TW/Ratio
Type
(Jerger types A, As, Ad, B, C, D and E denote the shape of the 226 Hz curve). You can set the type to be
determined automatically, and you can subsequently change it manually, if necessary.
E. On-screen selections
Curve selection
Manually adjusting the TPP
Changing tympanogram curve type
Deleting a curve
Otometrics - MADSEN Zodiac
SC
(Static Peak Compliance) when volume equivalent units are used
(Tympanometric Width/Tympanometric Ratio). Describes the steepness of the curve.
Press the
the Reflex Screening and Tympanometry result tables.
Press the
TPP, and press the
Press the
desired type is shown, and press the
Select the curve you wish to delete, and press the
To replace a single measurement, delete a curve and adjust the measurement
selections to redo the single measurement.
8 Testing with Zodiac Quick Check - Stand-Alone
Previous Curve
or
Next Curve
Edit TPP
softbutton, use the dial to move the marker to the desired
Done
softbutton.
Edit TPP
softbutton. Press the
Done
softbutton to select the desired curve in
Change Type
softbutton until the
softbutton.
Delete
button.