9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
Before testing
1. Do as described in
2. Inform the patient about the high stimulus levels in the test.
3. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
4. If contralateral stimulation is used in the test, make sure that both the ipsilateral probe and the contralateral phone
are in place before you start the measurement.
5. If needed, toggle the
Testing
Warning
ured results are kept.
Starting the test
1. Press the
Reflex Decay
2. Select
Reflex Decay
3. The
decay will light up on the buttons, e.g.
4. Use the dial to set the stimulus level. The level is shown in the gray on-
line values area of the display.
5. Prepare the patient for the high sound levels in the test. A measurement
takes about 10 seconds.
6. Ask the patient to sit very still and quiet during the test, without moving
head or jaw.
7. Press the
8. If needed, repeat these steps until you have collected the desired meas-
urements.
Warning
is interrupted immediately. Already measured results are kept.
64
Preparing for testing ► 36
Ear Selector
to select the ear on which you wish to start the test.
If the patient is troubled by the test, stop the test. The test is interrupted immediately. Already meas-
Special
button.
.
view is shown and the preselected settings for reflex
Start/Stop
button to start the test for the preset stimulus.
If the patient is troubled by the test, stop the test. The test
.
Contra
0.5 k
,
Hz.
Otometrics - MADSEN Zodiac