Reflex Decay graph
The reflex decay graph shows the reflex decay curve and the half-life time for the selected stimulus.
Time line in seconds
Stimulus on/stimulus off markers
Deflection curve for the selected stimulus
Half-life marker
Reflex Decay results
Use the buttons in the Reflex section on the front panel to navigate between the measurements. Your selection is high-
lighted in the Reflex results table.
The results table shows the results related to the currently selected curve:
Stimulus side
Stimulus frequency
Half-life time
If the system cannot automatically analyze a reflex measurement, a
On-screen selections
Adjust Marker
Hide Marker
Done
10

Managing test results in MADSEN Zodiac

The following applies to managing test results in MADSEN Zodiac. If you wish to perform these actions in OTOsuite, see
the OTOsuite User Guide or OTOsuite Reference Manual.
Swapping ear results
If you have tested a patient with the wrong ear selected, you can swap the ear results both during a session and after a ses-
sion so that the results are assigned to the other ear. Swapping applies to all tests made on that patient.
Deleting measurements
1. You can delete a single measurement curve by using the soft-button functions
the curve you wish to delete.
Otometrics - MADSEN Zodiac
Press this soft-button so that you can adjust the half-life time marker by using the
dial.
/
Show Marker
If, for example, the result is ambiguous, you can hide a half-life time marker.
Alternatively, you can show the marker.
When you have adjusted the marker, press the
The results table is updated with the selected value.
Not Tested
indication is shown.
Done
soft-button.
Previous Curve
or
User Guide
Next Curve
to select
33