9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
A. On-line values
The on-line values show the current settings and status.
Probe tone in Hz
Pump speed
Sweep direction
Current equivalent volume/admittance (678, 800, 1000 Hz probe tone)
B. Tympanometry graph
The graph area shows the tympanometric curves and can rescale automatically to fit the curves.
Tympanometric curves
Pressure and admittance scales
Ear canal volume bar. The ear canal volume is shown to the right of the graph.
Norm area
C. Reflex Screening results
The detected reflex is shown in the results area.
Stimulus type level
Progress indicator during measurements
Deflection curves are shown when a reflex has been detected.
If a reflex is not detected, the highest stimulus is displayed together with a
If the system cannot automatically analyze a reflex measurement, a
D. Tympanometry results
The results table shows the results related to the currently selected curve:
Probe
(the probe tone in Hz)
TPP
(Tympanometric Peak Pressure)
ECV
(Equivalent Ear Canal Volume)
SA
(Static Peak Admittance), or
TW/Ratio
Type
(Jerger types A, As, Ad, B, C, D and E denote the shape of the 226 Hz curve). You can set the type to be
determined automatically, and you can subsequently change it manually, if necessary.
56
SC
(Static Peak Compliance) when volume equivalent units are used
(Tympanometric Width/Tympanometric Ratio). Describes the steepness of the curve.
No Response
indication.
Not Tested
indication is shown.
Otometrics - MADSEN Zodiac