12 Testing with Zodiac Diagnostic or Clinical - PC-based
A. Reflex Decay graph
The graph shows the half-life time and the reflex decay curve for the selected stimulus.
•
Stimulus on-set marker (if needed, click and drag to adjust)
•
Half-life marker (if needed, click and drag to adjust)
•
Time line in seconds
•
Stimulus on/stimulus off markers
•
Deflection curve for the selected stimulus
B. Live baseline graph
The live baseline graph starts measuring the baseline as soon as a successful probe fit is achieved. It illustrates the sta-
bility of the physical measurement conditions. The stability is reflected directly in the deflection curves if a meas-
urement is made.
With particularly steep tympanograms, the stability of the baseline can often be improved by off-setting the target pres-
sure from the tympanometric peak pressure.
Variations that are in synchrony with the patient's pulse or breathing may prevent the measurements from being suc-
cessful.You can use the Admittance Recording test to record and report such variations (Zodiac Clinical only).
C. Tympanogram
The primary tympanogram is shown in this view.
Decay data table
You can click the individual measurements in the data table to view and edit the corresponding curves.
This table shows the settings and measurement results.
•
Probe tone in Hz
•
Stimulus side
•
Stimulus frequency
•
Stimulus on/stimulus off markers
Editing results
If needed, click and drag to adjust
Select the curve in the results
table and right-click the curve to
edit the properties.
100
•
The stimulus on-set marker
•
The half-life marker
•
Delete Curve
Swap Data Between Left and Right Ear
•
•
Delete Marker
Add Marker
•
Otometrics - MADSEN Zodiac