1. To remove the probe tip, hold the probe by
the probe body and twist the probe ring
slightly counter-clockwise. This will loosen the
probe tip.
Caution
•
ture may dissolve any residual cerumen and
thus contaminate the sensitive parts in the
body of the probe.
2. Take out the probe tip.
3. Fit the probe tip on the probe, and twist the
probe ring clockwise to lock the probe tip in
place on the probe.
Probe check
If you have fitted a new probe tip, do a probe check. See
17.4.6
Cleaning the test cavities
If a test cavity becomes contaminated, do not use it. Dispose of it and replace it with a new one.
17.4.7
Cleaning the printer
Check the paper compartment regularly to make sure that it is free from paper dust and that the thermal paper head is
clean.
Prerequisites
•
Before cleaning, switch off MADSEN Zodiac and disconnect it from any external power source.
Warning
•
Caution
•
Otometrics - MADSEN Zodiac
Even the slightest amount of mois-
When you open the paper compartment, be careful not to injure your hands on the cutter blades.
When you clean the paper compartment, never use metal objects or liquids.
Probe check ► 36
.
17 Service and maintenance