12.6.2

Automated Reflex Threshold testing

In the following you will find the description for performing automated reflex threshold testing. The automated search for
the reflex threshold can be used both for individual reflex stimulus types and for preconfigured sequence testing.
The following procedure is an example describing how you search for a single stimulus type threshold.
Starting the test
1. Select
Auto
2. Ask the patient to sit very still and quiet during the test, without moving
head or jaw.
3. Prepare the patient for the high sound levels in the test.
4. Click either the
5. Select the frequency that you wish to test: Click the applicable arrow but-
ton.
6. Check the baseline graph to make sure that you have a stable test con-
dition before you start presenting the stimulus. The baseline indicator
should be as close to zero as possible. If needed, you can offset the pres-
sure to stabilize your measurements: Click the applicable arrow button.
7. Make sure that the settings options for the automated reflex sequence
have been selected. They include starting level, increment size, veri-
fication procedure, and maximum level.
As opposed to a full measurement sequence, the automated threshold
search only includes a single, selected stimulus side and stimulus type.
Note
You will be warned about a high stimulus intensity level if you
reach the warning limit.
8. Click the
tion curve is plotted in real time.
9. If needed, repeat these steps until you have collected the desired meas-
urements for the stimulus sides and types you wish to test.
Warning
is interrupted immediately. Already measured results are kept.
The search will be paused before each stimulus above 100 dB HL, and you will
be prompted to continue.
Stopping the test
The test stops automatically, but can be stopped manually, if needed.
1. Click the
2. The test is interrupted immediately.
Otometrics - MADSEN Zodiac
testing.
Ipsi
Contra
reflex or the
Start
button to start the automated threshold search. The deflec-
If the patient is troubled by the test, stop the test. The test
Stop
button on the Control Panel or the
12 Testing with Zodiac Diagnostic or Clinical - PC-based
reflex button.
Start/Stop
button on the probe.
The baseline indicator starts imme-
diately when you insert the probe in
the ear of the patient.
95