17 Service and maintenance
If there is a probe error
In case of a probe error, check the probe status for
possible causes.
If the probe is faulty, contact an authorized service
department for repair.
17.4

Cleaning MADSEN Zodiac

Caution
Cleaning agents
Caution
See
Recommended cleaning agents ► 119
Frequency
We recommend that you set up a schedule for cleaning Zodiac and accessories such as probes and/or earphones.
Prerequisites
Before cleaning, switch off MADSEN Zodiac and disconnect it from any external power source.
If needed, unplug the contralateral probe from MADSEN Zodiac.
Cleaning under the dial
If the device is a Zodiac Diagnostic or a Zodiac Clinical, use a thin cloth with a recommended cleaning agent to gently
the clean the area beneath it.
If needed, use a screw driver (torq IP6) to carefully unscrew the screw fastening the dial.
Procedure
Warning
damage the instrument and it may result in a risk of electrical shock to the user or patient.
1. Clean the external surfaces of the device:
Remove dust using a soft brush.
Use a soft, slightly damp cloth with a small amount of cleaning agent to clean the external surfaces of the device
according to local infection control regulations.
See the list of recommended cleaning agents.
2. Make sure that you have cleaned all surfaces, edges and crevices thoroughly.
3. Reconnect the contralateral probe, if needed.
118
Make sure that you comply with local infection control regulations.
Use only the cleaning agents prescribed for cleaning the device.
Keep the unit away from liquids. Do not allow moisture inside the unit. Moisture inside the unit can
.