• For a manual trace measurement, use the 
touchpad to position the caliper. Press the 
 key. Use the touchpad to trace the 
SELECT
waveform.
If you need to make a correction, select 
Undo on‐screen or press the 
key.
• For an auto trace measurement, select 
Auto on‐screen, and use the touchpad to 
position the first caliper at the beginning 
of the waveform. Press the 
position the second caliper at the end of 
the waveform.
Confirm that the system‐generated 
boundary is correct. If you are not satisfied 
with the trace, obtain a higher quality 
Doppler spectral trace image, or trace 
manually.
c Press the 
 key.
SET
d
Save the calculation. (See "To save a 
calculation" on page 51.)

Vascular calculations

WARNING:
 
BACKSPACE
 key, and 
SELECT
Transducer
C11x
HFL38x
L25x
L38x
L38xi
SLAx
The vascular measurements that you can save to 
the patient report are listed in the following table. 
For definitions of acronyms, see "Glossary" on 
page 189
To avoid misdiagnosis or harming
the patient outcome, start a new
patient information form before
starting a new patient exam and
performing calculations. Starting a
new patient information form
clears the previous patient's data.
The previous patient's data will be
combined with the current patient
if the form is not first cleared. See
"To create a new patient
information form"
To avoid incorrect calculations,
verify that the patient information,
date, and time settings are
accurate.
Exam Type
Vascular
Vascular
Vascular
Vascular
Vascular
Vascular
Chapter 4: Measurements and Calculations
on page 40.
75