Advertisement

5 Positioning the calipers, measure the 
ventricular length. (See "Working with 
calipers" on page 47.)
6 Save the calculation.
To measure peak velocity
For each cardiac measurement, the system saves 
up to five individual measurements and 
calculates their average. If you take more than 
five measurements, the most recent measurement 
replaces the fifth one. If you delete a saved 
measurement from the patient report, the next 
measurement taken replaces the deleted one in 
the patient report. The most recently saved 
measurement appears at the bottom of the 
calculations menu. 
1 On a frozen Doppler spectral trace, press the 
 key. 
CALCS
2 From the calculations menu, select MV, TV, 
TDI, or P. Vein.
3 Do the following for each measurement you 
want to take:
a Select the measurement name from the 
calculations menu. 
b
Position the calipers. (See "Working with 
calipers" on page 47.)
c
Save the calculation. (See "To save a 
calculation" on page 51.)
To calculate Velocity Time Integral (VTI)
Note: This calculation computes other results in 
addition to VTI. See the table "Cardiac Calculations" 
on page 57.
1 On a frozen Doppler spectral trace, press the 
 key.
CALCS
2 From the calculations menu, select VTI under 
MV, AV, TV, PV, or LVOT.
3 Position the caliper at the start of the 
waveform, and press the 
the trace.
 key to start 
SELECT
4 Using the touchpad, trace the waveform.
To make a correction, select Undo on‐screen, 
backtrack with the touchpad, or press the 
 key.
BACKSPACE
5 Press the 
 key to complete the trace.
SET
Save the calculation. (See "To save a 
6
calculation" on page 51.)
For information on the automatic trace tool, see 
"To trace automatically (Doppler)" on page 50.
To calculate Right Ventricular Systolic
Pressure (RVSP)
1 On a frozen Doppler spectral trace, press the 
 key.
CALCS
2 From the calculations menu, select TV and 
then select TRmax.
3
Position the caliper. (See "Working with 
calipers" on page 47.)
4
Save the calculation. (See "To save a 
calculation" on page 51.)
To adjust the RA pressure, see "To delete a 
5
vascular or cardiac measurement" on page 77.
Changing the RA pressure from the default 5 
affects the RVSP calculation in the patient 
report.
To calculate Pressure Half Time (PHT) in MV,
AI, or TV
1 On a frozen Doppler spectral trace, press the 
 key.
CALCS
2 From the calculations menu, select MV, AV, or 
TV, and then select PHT.
3 Position the first caliper at the peak, and press 
the 
 key.
SELECT
A second caliper appears.
4 Position the second caliper:
• In MV, position the caliper along the EF 
slope.
Chapter 4: Measurements and Calculations
61