Cleaning and disinfecting the battery

Caution:
To avoid damaging the battery, do
not allow cleaning solution or
disinfectant to come in contact
with the battery terminals.
To clean and disinfect a battery (wipe
method)
1 Remove the battery from the system.
2 Clean the surface using a soft cloth lightly 
dampened in a mild soap or detergent 
cleaning solution.
Apply the solution to the cloth rather than the 
surface.
3 Wipe the surfaces with the disinfection 
solution. Sani‐Cloth HB, Sani‐Cloth Wipes, or 
70% isopropyl alcohol is recommended.
4 Air dry.

Cleaning the footswitch

Caution:
To avoid damaging the footswitch,
do not sterilize. It is not intended for
use in a sterile environment.
To clean the footswitch
1 Dampen a non‐abrasive cloth with one of the 
following products:
• Isopropyl alcohol
• Soap and water
• Cidex
• Sodium Hypochlorite 5.25% (Bleach) 
diluted 10:1
2 Wring out cloth until slightly wet and then 
gently rub soiled area until clean.
86
Maintenance

Cleaning and disinfecting ECG cables

Caution:
To avoid damaging the ECG cable,
do not sterilize.
To clean and disinfect the ECG cable (wipe
method)
1 Remove the cable from the system.
2 Clean the surface using a soft cloth lightly 
dampened in a mild soap or detergent 
cleaning solution.
Apply the solution to the cloth rather than the 
surface.
3 Wipe the surfaces with any of the following 
products:
• Bleach (sodium hypochlorite)
• Cidex disinfectants
• Green soap
4 Air dry or towel dry with a clean cloth.