Presets Setup - FujiFilm SonoSite M-Turbo User Manual

Ultrasound system
Hide thumbs Also See for SonoSite M-Turbo:
Table of Contents

Advertisement

Gestational Age Table Measurements The system 
provides gestational age measurements by 
selected authors for GS, CRL, BPD, OFD, HC, 
TTD, APTD, AC, FTA, FL, EFW, Tibia, HL, and 
5 additional custom measurement labels.
Growth Analysis Table Measurements The 
system provides growth graphs or curves for 
BPD, HC, AC, FL, EFW, and HC/AC.
WARNING:
Prior to use, verify that custom table
data entries are correct. The system
does not confirm the accuracy of
the custom table data entered by
the user.
To view OB tables
1 On the OB Calculations or OB Custom 
Measurements setup page, select Tables 
on‐screen.
2 Select the desired table and 
measurement/author.
To create a new OB custom table
You can create two custom tables for each OB 
measurement. 
1 On the OB Calculations or OB Custom 
Measurements setup page, select Tables 
on‐screen.
2 Select the desired table (Gestational Age or 
Growth Analysis).
3 In the Measurement list, select the 
measurement for the custom table.
4 Select New on‐screen.
5 In the Author box, type a unique name.
6 Enter the data.
7 Select Save on‐screen.
To display the measurement for the custom table 
in the calculations menu, see "To specify 
gestational age and growth analysis" on page 20.
22

Presets setup

To edit or delete an OB custom table
1 On the OB Calculations or OB Custom 
Measurements setup page, select Tables 
on‐screen.
2 Select the OB custom table.
3 Select one of the following on‐screen:
• Edit Enter data, and then select Save 
on‐screen.
• Delete to remove the custom table. Select 
Yes.
Presets setup
The Presets setup page has settings for general 
preferences. You can select from the following 
lists:
Doppler Scale Select cm/s or kHz.
Duplex The layout for displaying M Mode trace 
and Doppler spectral trace:1/3 2D, 2/3 Trace; 1/2
2D, 1/2 Trace; or Full 2D, Full Trace.
Live Trace Select Peak or Mean.
Thermal Index You can select TIS, TIB, or TIC. The 
default setting is based on exam type: OB is TIB, 
TCD is TIC, and all others are TIS. 
Save Key Behavior of the 
saves the image to internal storage. Image/Calcs 
saves the image to internal storage and saves the 
current calculation to the patient report.
Dynamic Range Settings include -3, -2, -1, 0, +1, 
+2, or +3. Negative numbers show higher
contrast images, and positive numbers show 
lower contrast images.
Units Units for patient height and weight in 
cardiac exams: in/ft/lbs or cm/m/kg.
Language The system language. Changing the 
language requires restarting the system.
Color Scheme The background color of the 
display.
 key. Image Only 
SAVE

Hide quick links:

Advertisement

Table of Contents
loading