Gestational Age Table Measurements The system
provides gestational age measurements by
selected authors for GS, CRL, BPD, OFD, HC,
TTD, APTD, AC, FTA, FL, EFW, Tibia, HL, and
5 additional custom measurement labels.
Growth Analysis Table Measurements The
system provides growth graphs or curves for
BPD, HC, AC, FL, EFW, and HC/AC.
WARNING:
Prior to use, verify that custom table
data entries are correct. The system
does not confirm the accuracy of
the custom table data entered by
the user.
To view OB tables
1 On the OB Calculations or OB Custom
Measurements setup page, select Tables
on‐screen.
2 Select the desired table and
measurement/author.
To create a new OB custom table
You can create two custom tables for each OB
measurement.
1 On the OB Calculations or OB Custom
Measurements setup page, select Tables
on‐screen.
2 Select the desired table (Gestational Age or
Growth Analysis).
3 In the Measurement list, select the
measurement for the custom table.
4 Select New on‐screen.
5 In the Author box, type a unique name.
6 Enter the data.
7 Select Save on‐screen.
To display the measurement for the custom table
in the calculations menu, see "To specify
gestational age and growth analysis" on page 20.
22
Presets setup
To edit or delete an OB custom table
1 On the OB Calculations or OB Custom
Measurements setup page, select Tables
on‐screen.
2 Select the OB custom table.
3 Select one of the following on‐screen:
• Edit Enter data, and then select Save
on‐screen.
• Delete to remove the custom table. Select
Yes.
Presets setup
The Presets setup page has settings for general
preferences. You can select from the following
lists:
Doppler Scale Select cm/s or kHz.
Duplex The layout for displaying M Mode trace
and Doppler spectral trace:1/3 2D, 2/3 Trace; 1/2
2D, 1/2 Trace; or Full 2D, Full Trace.
Live Trace Select Peak or Mean.
Thermal Index You can select TIS, TIB, or TIC. The
default setting is based on exam type: OB is TIB,
TCD is TIC, and all others are TIS.
Save Key Behavior of the
saves the image to internal storage. Image/Calcs
saves the image to internal storage and saves the
current calculation to the patient report.
Dynamic Range Settings include -3, -2, -1, 0, +1,
+2, or +3. Negative numbers show higher
contrast images, and positive numbers show
lower contrast images.
Units Units for patient height and weight in
cardiac exams: in/ft/lbs or cm/m/kg.
Language The system language. Changing the
language requires restarting the system.
Color Scheme The background color of the
display.
key. Image Only
SAVE