Reexport the original data onto the USB storage
device.
Contact your system administrator.
A maintenance icon
screen. Restart the system. If the issue recurs,
system maintenance may be required. Note the
number that appears in parentheses on the C: line
and contact SonoSite or your SonoSite
representative.Software licensing
SonoSite software is controlled by a license key.
After you install new software, the system
prompts you for a license key. You must obtain
one key for each system or transducer that uses
the software.
The software will operate for a short time (the
grace period) without a license key. During the
grace period, all system functions are available.
After the grace period, the system is not usable
until you enter a valid license key. Grace period
time is not used while the system is off or asleep.
Grace period time remaining appears on the
license update screen.
Caution:
After the grace period expires, all
system functions except licensing
are unavailable until a valid license
key is entered.
To obtain a license key for your software, contact
SonoSite Technical Support. (See "SonoSite
Technical Support" on page 7.) You need to
provide the following information. (See "System
Information setup" on page 23.)
System Software
Name of person
installing the
upgrade
Serial number (on
bottom of system)
82
Maintenance
appears on the system
Transducer Software
Name of person installing
the upgrade
Transducer serial number
System Software
ARM version
PCBA serial
number
After you obtain a license key, you must enter it
into the system.
To enter a license key
1 Turn on the system.
The license update screen appears.
2 Enter the license key in the Enter license
number field.
3 Select Done on‐screen.
If you entered a valid license key but the
license update screen appears, verify that you
entered the license key correctly. If the license
update screen still appears, contact SonoSite
Technical Support. (See "SonoSite Technical
Support" on page 7.)
Maintenance
Use the recommendations in this section when
cleaning or disinfecting your ultrasound system,
transducer, and accessories. Use the cleaning
recommendations in the peripheral
manufacturer's instructions when cleaning or
disinfecting your peripherals.
No periodic or preventive maintenance is
required for the system, transducer, or
accessories other than cleaning and disinfecting
the transducer after every use. (See "Cleaning
and disinfecting transducers" on page 84.) There
are no internal components that require periodic
testing or calibration. All maintenance
requirements are described in this chapter and in
the ultrasound system service manual.
Transducer Software
Transducer part number
(REF)
or model number (for
example, C60x)
Transducer bundle version