Reexport the original data onto the USB storage
device.
Contact your system administrator.
A maintenance icon
screen. Restart the system. If the issue recurs, 
system maintenance may be required. Note the 
number that appears in parentheses on the C: line 
and contact SonoSite or your SonoSite 
representative.Software licensing
SonoSite software is controlled by a license key. 
After you install new software, the system 
prompts you for a license key. You must obtain 
one key for each system or transducer that uses 
the software.
The software will operate for a short time (the 
grace period) without a license key. During the 
grace period, all system functions are available. 
After the grace period, the system is not usable 
until you enter a valid license key. Grace period 
time is not used while the system is off or asleep. 
Grace period time remaining appears on the 
license update screen. 
Caution:
After the grace period expires, all
system functions except licensing
are unavailable until a valid license
key is entered.
To obtain a license key for your software, contact 
SonoSite Technical Support. (See "SonoSite 
Technical Support" on page 7.) You need to 
provide the following information. (See "System 
Information setup" on page 23.)
System Software
Name of person
installing the
upgrade
Serial number (on
bottom of system)
82

Maintenance

appears on the system
Transducer Software
Name of person installing
the upgrade
Transducer serial number
System Software
ARM version
PCBA serial
number
After you obtain a license key, you must enter it 
into the system.
To enter a license key
1 Turn on the system.
The license update screen appears.
2 Enter the license key in the Enter license
number field.
3 Select Done on‐screen.
If you entered a valid license key but the 
license update screen appears, verify that you 
entered the license key correctly. If the license 
update screen still appears, contact SonoSite 
Technical Support. (See "SonoSite Technical 
Support" on page 7.)
Maintenance
Use the recommendations in this section when 
cleaning or disinfecting your ultrasound system, 
transducer, and accessories. Use the cleaning 
recommendations in the peripheral 
manufacturer's instructions when cleaning or 
disinfecting your peripherals.
No periodic or preventive maintenance is 
required for the system, transducer, or 
accessories other than cleaning and disinfecting 
the transducer after every use. (See "Cleaning 
and disinfecting transducers" on page 84.) There 
are no internal components that require periodic 
testing or calibration. All maintenance 
requirements are described in this chapter and in 
the ultrasound system service manual. 