To trace IMT manually
In manually tracing IMT, the user defines the 
location.
1 On a frozen 2D image, press the 
2 From the calculations menu, select a 
measurement name.
3 Select Edit on‐screen, and then select Manual, 
and then select Sketch.
A single caliper appears, and Trace appears 
next to the measurement.
4 Do the following for the desired 
adventitia‐media boundary and then for the 
lumen‐intima boundary:
a Position the caliper at the beginning of the 
boundary, and press the 
b Using the touchpad, mark points by 
moving the caliper to the next desired 
point and pressing the 
To make a correction, select Undo 
on‐screen or press the 
delete the last segment.
c Press the 
 key to complete the trace line.
SET
Save the calculation. (See "To save a 
5
calculation" on page 51.)
To sketch IMT
The IMT sketch measurement involves two 
user‐defined sketch lines that you can adjust 
manually.
1 On a frozen 2D image, press the 
2 From the calculations menu, select a 
measurement name.
3 Select Edit on‐screen, and then select Manual.
A single caliper appears on‐screen, and Sketch 
appears next to the measurement.
4 Do the following for the desired 
adventitia‐media boundary and then for the 
lumen‐intima boundary:
 key
CALCS
 key.
SELECT
 key.
SELECT
 key to 
BACKSPACE
 key
CALCS
a Position the caliper at the beginning of the 
boundary and press the 
b Using the touchpad, mark points by 
moving the caliper to the next desired 
point and pressing the 
To make a correction, select Undo 
on‐screen or press the 
delete the last segment.
c Press the 
 key to complete the trace line.
SET
d If necessary, adjust or edit the 
measurement. See "IMT tool options" on 
page 68.
e
Save the calculation. (See "To save a 
calculation" on page 51.)

OB calculations

EFW is calculated only after appropriate 
measurements are completed. If any one of these 
parameters results in an EDD greater than what 
the OB calculation tables provide, the EFW is not 
displayed.
WARNING:
Make sure that you have selected
the OB exam type and the OB
author for the OB calculation table
you intend to use. See
System-Defined OB Measurements
and Table Authors"
To avoid incorrect obstetrics
calculations, verify with a local clock
and calendar that the system's date
and time settings are correct before
each use of the system. The system
does not automatically adjust for
daylight savings time changes.