FujiFilm SonoSite M-Turbo User Manual page 67

Ultrasound system
Hide thumbs Also See for SonoSite M-Turbo:
Table of Contents

Advertisement

Transducer
Exam Types
C11x
Abdomen, Vascular
C60x
Abdomen
HFL38x
Vascular
L25x
Vascular
L38x
Vascular
L38xi
Vascular
P10x
Abdomen
P21x
Abdomen
SLAx
Vascular
The following table shows the measurements 
required to complete the volume flow 
calculation. For definitions of acronyms, see 
"Glossary" on page 189.
Volume Flow Calculations
Menu
Measurement
Heading
(Imaging Mode)
Vol Flow
D (2D)*
TAM or TAP
(Doppler)
* Required if measuring the diameter instead of using the gate
size
Both a 2D and a Doppler measurement are 
required for the volume flow calculation. For the 
2D measurement, you can do either of the 
following:
• Measure the diameter of the vessel. This 
approach is more precise. The measurement 
overrides the gate size. 
• Use the gate size. If you do not measure the 
diameter of the vessel, the system 
The Doppler sample volume should completely 
insonate the vessel. You can measure either the 
time average mean or time average peak. To 
specify the live trace setting, see "Presets setup" 
on page 22.
Consider the following factors when performing 
volume flow measurements:
• Users should follow current medical practice 
• The accuracy of the volume flow calculation 
• The factors identified in the literature that 
Calculation
Result
VF (Volume
Flow ml/min)
The considerations and degree of accuracy for 
volume flow measurements and calculations are 
discussed in the following references:
automatically uses the gate size and "(gate)" 
appears in the calculation results. 
for volume flow calculation applications.
largely depends on the user.
affect the accuracy are as follows:
• Using the diameter method for 2D area
• Difficulty ensuring uniform insonation of 
the vessel. 
The system is limited to the following 
sample volume sizes:
• C11x transducer: 1, 2, 3 Gate Size (mm)
• C60x and P10x transducers: 2, 3, 5, 7, 
10, 12 Gate Size (mm)
• HFL38x, L25x, and SLAx transducers: 
1, 3, 5, 6, 7, 8, 10, 12 Gate Size (mm)
• L38x transducer: 1, 3, 5, 7, 10, 12 Gate 
Size (mm)
• P21x transducer: 2, 3, 5, 7, 11.5, 14 Gate 
Size (mm)
• Precision in placing the caliper
• Accuracy in angle correction
Allan, Paul L. et al. Clinical Doppler Ultrasound, 
4th Ed., Harcourt Publishers Limited, (2000) 
36‐38.
Chapter 4: Measurements and Calculations
55