Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The M-Turbo ultrasound system is intended for use in the electromagnetic environment specified
below. The customer or the user of the M-Turbo ultrasound system should assure that it is used in
such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
Transient burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
110
IEC 60601 Test Level
±6.0KV contact
±8.0KV air
±2KV for power
supply lines
±1KV for input/output
lines
±1KV line(s) to line(s)
±2KV line(s) to earth
<5% U
T
(>95% dip in U
) for
T
0.5 cycle
40% U
T
(60% dip in U
) for 5
T
cycles
70% U
T
(30% dip in U
) for 25
T
cycles
<5% U
T
(>95% dip in U
) for 5s
T
Electromagnetic
Compliance Level
Environment
±6.0KV contact
Floors should be wood,
concrete, or ceramic tile. If
±8.0KV air
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
±2KV for power
Mains power quality should
supply lines
be that of a typical
commercial or hospital
±1KV for input/
environment.
output lines
±1KV line(s) to
Mains power quality should
line(s)
be that of a typical
commercial or hospital
±2KV line(s) to
environment.
earth
<5% U
Mains power quality should
T
be that of a typical
(>95% dip in U
)
T
commercial or hospital
for 0.5 cycle
environment. If the user of the
40% U
T
M-Turbo ultrasound system
(60% dip in U
) for
requires continued operation
T
during power mains
5 cycles
interruptions, it is
70% U
T
recommended that the
(30% dip in U
) for
T
M-Turbo ultrasound system be
25 cycles
powered from an
<5% U
uninterruptible power supply
T
or a battery.
(>95% dip in U
)
T
for 5s