floating‐point versus integer‐type math, which is 
subject to errors introduced by rounding versus 
truncating results for display of a given level of 
significant digit in the calculation.
Measurement publications
and terminology
The following sections list the publications and 
terminology used for each calculation result.
Terminology and measurements comply with 
AIUM published standards.

Cardiac references

Acceleration (ACC) in cm/s
Zwiebel, W.J. Introduction to Vascular 
Ultrasonography. 4th ed., W.B. Saunders 
Company, (2000), 52.
ACC = abs (delta velocity/delta time)
Acceleration Time (AT) in msec
Oh, J.K., J.B. Seward, A.J. Tajik. The Echo Manual. 
2nd ed., Lippincott, Williams, and Wilkins, 
(1999), 219.
Aortic Valve Area (AVA) by Continuity
2
Equation in cm
Reynolds, Terry. The Echocardiographer's Pocket 
Reference. 2nd ed., School of Cardiac Ultrasound, 
Arizona Heart Institute, (2000), 393, 442.
A
 = A
 * V
/V
2
1
1
2
where:
A
 = Ao valve area
2
A
 = LVOT area; V
1
V
 = Ao valve velocity
2
LVOT = Left Ventricular Outflow 
Tract
AVA (PV
/PV
) * CSA
LVOT
AO
AVA (VTI
/VTI
) * CSA
LVOT
AO
2
 = LVOT velocity; 
1
LVOT
LVOT
Body Surface Area (BSA) in m
Grossman, W. Cardiac Catheterization and 
Angiography. Philadelphia: Lea and Febiger, 
(1980), 90.
BSA = 0.007184 * Weight
Weight = kilograms
Height = centimeters
Cardiac Index (CI) in l/min/m
Oh, J.K., J.B. Seward, A.J. Tajik. The Echo Manual. 
2nd Edition, Boston: Little, Brown and Company, 
(1999), 59.
CI = CO/BSA
where:
CO = Cardiac Output
BSA = Body Surface Area
Cardiac Output (CO) in l/min
Oh, J.K., J.B. Seward, A.J. Tajik The Echo Manual. 
2nd ed., Lippincott, Williams, and Wilkins, 
(1999), 59.
CO = (SV * HR)/1000
where:
CO = Cardiac Output
SV = Stroke Volume
HR = Heart Rate
Cross Sectional Area (CSA) in cm
Reynolds, Terry. The Echocardiographer's Pocket 
Reference. 2nd ed., School of Cardiac Ultrasound, 
Arizona Heart Institute, (2000), 383.
CSA = 0.785 * D
2
where:
D = diameter of the anatomy of 
interest
Deceleration Time in msec
Reynolds, Terry. The Echocardiographer's Pocket 
Reference. 2nd ed., School of Cardiac Ultrasound, 
Arizona Heart Institute, (2000), 453.
|time a ‐ time b|
Chapter 7: References
2
0.425
 * Height
0.725
2
2
173