FujiFilm SonoSite M-Turbo User Manual page 88

Ultrasound system
Hide thumbs Also See for SonoSite M-Turbo:
Table of Contents

Advertisement

.
Vascular Calculations
Menu
Vascular
Heading
Measurement
CCA
Prox
Mid
Dist
Bulb
ICA
Prox
Mid
Dist
ECA
Prox
Mid
Dist
VArty
To perform a Vascular calculation
After you perform vascular measurements, 
values in the ICA/CCA ratios are selectable on the 
vascular page of the patient report.
1 On a frozen Doppler spectral trace, press the 
 key. 
CALCS
2 From the calculations menu, select Left or 
Right.
76
Exam-based calculations
Calculation
Results
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
s (systolic),
d (diastolic)
3 Do the following for each measurement you 
want to take:
a From the calculations menu, select the 
measurement name.
b Using the touchpad, position the caliper at 
the peak systolic waveform.
c Press the 
SELECT
A second caliper appears.
d Using the touchpad, position the second 
caliper at the end diastole on the 
waveform.
e
Save the calculation. (See "To save a 
calculation" on page 51.)
 key.