4 Press the 
 key.
SELECT
5 Using the touchpad, complete the trace, and 
press the 
 key.
SET
See "To save a measurement to a calculation and 
patient report" on page 47.

M Mode measurements

The basic measurements that you can perform in 
M Mode imaging are as follows:
• Distance in cm/Time in seconds
• Heart Rate (HR) in beats per minute (bpm)
The time scale above the trace has small marks at 
200 ms intervals and large marks at one‐second 
intervals.
To measure distance (M Mode)
You can perform up to four distance 
measurements on an image.
1 On a frozen M Mode trace, press the 
key.
A single caliper appears.
2 Using the touchpad, position the caliper.
3 Press the 
 key to display the second 
SELECT
caliper.
4 Using the touchpad, position the second 
caliper.
See "To save a measurement to a calculation and 
patient report" on page 47.
To measure heart rate (M Mode)
1 On a frozen M Mode trace, press the 
key.
2 Select HR on‐screen.
A vertical caliper appears.
3 Using the touchpad, position the vertical 
caliper at the peak of the heartbeat. 
4 Press the 
A second vertical caliper appears. 
5 Using the touchpad, position the second 
vertical caliper at the peak of the next 
heartbeat.
See "To save a measurement to a calculation and 
patient report" on page 47. Saving the heart rate 
measurement to the patient report overwrites 
any heart rate entered on the patient information 
form.
See also "To measure fetal heart rate (M Mode)" 
on page 71.

Doppler measurements

The basic measurements that you can perform in 
Doppler imaging are Velocity (cm/s), Pressure 
Gradient, Elapsed Time, +/x Ratio, Resistive 
Index (RI), and Acceleration. You can also trace 
manually or automatically. 
 
CALIPER
For Doppler measurements, the Doppler scale 
must be set to cm/s. See "Presets setup" on 
page 22.
To measure Velocity (cm/s) and Pressure
Gradient (Doppler)
1 On a frozen Doppler spectral trace, press the 
CALIPER
A single caliper appears.
2 Using the touchpad, position the caliper to a 
peak velocity waveform.
This measurement involves a single caliper from 
the baseline.
 
CALIPER
See "To save a measurement to a calculation and 
patient report" on page 47.
Chapter 4: Measurements and Calculations
 key.
SELECT
 key.
49