3 Using the touchpad, position the vertical 
caliper at the peak of the heartbeat. 
4 Press the 
 key.
SELECT
A second vertical caliper appears. 
5 Using the touchpad, position the second 
vertical caliper at the peak of the next 
heartbeat.
Save the calculation. (See "To save a 
6
calculation" on page 51.)
OB Doppler Calculations
Menu
OB
Heading
Calculation
MCA (Middle
S/D, RI
Cerebral
Artery)
S/D, RI, PI*
Umb A
S/D, RI
(Umbilical
Artery)
S/D, RI, PI*
*Calculation requires a trace measurement.
To calculate MCA or Umba (Doppler)
Note: The system does not provide an MCA/UmbA 
ratio from the PI (Pulsatility Index). 
1 Select OB exam type, and select LMP or 
Estab.DD in the patient information form.
2 On a frozen Doppler spectral trace, press the 
 key.
CALCS
3 Do the following for each measurement you 
need to take:
a From the calculations menu, select the 
measurement name under MCA (Middle 
72
Exam-based calculations
Results
SD
RI
SD
RI
PI

Small Parts calculations

SD
Small Parts calculations include volume, hip 
RI
angle, and d:D ratio. For instructions to calculate 
volume, see "Volume calculations" on page 54. 
SD
Transducer
RI
PI
HFL38x
HFL50x
L38x
L38xi
To calculate hip angle
1 On a frozen 2D image, press the 
2 From the calculations menu, select Right or 
3 Select Baseline under Hip Angle.
Cerebral Artery) or UmbA (Umbilical 
Artery).
b Position the calipers:
• For S/D, RI, position the first caliper at 
the peak systolic waveform. Press the 
 key, and position the second 
SELECT
caliper at the end diastole on the 
waveform.
• For S/D, RI, PI, position the caliper at 
the beginning of the desired waveform, 
and press the 
SELECT
touchpad to manually trace the desired 
area. Press the 
SET
If calipers are not positioned correctly, 
the calculation result is inaccurate.
Save the calculation. (See "To save a 
c
calculation" on page 51.) 
Only one calculation (S/D, RI or S/D, RI, PI) 
can be saved.
Exam Type
Small Parts
Small Parts
Small Parts
Small Parts
Left.
A baseline appears on‐screen.
 key. Use the 
 key.
 key.
CALCS

Hide quick links: