FujiFilm SonoSite M-Turbo User Manual page 75

Ultrasound system
Hide thumbs Also See for SonoSite M-Turbo:
Table of Contents

Advertisement

4 Press the 
 key.
SELECT
A second horizontal dotted line with an active 
caliper appears at 300 cm/s.
5 Position the second caliper along the 
waveform at 300 cm/s.
Save the calculation. (See "To save a 
6
calculation" on page 51.)
To calculate Aortic Valve Area (AVA)
The AVA calculation requires a measurement in 
2D and two measurements in Doppler. After the 
measurements are saved, the result appears in 
the patient report.
1 Measure from LVOT (2D):
a On a frozen 2D image, press the 
b From the calculations menu, select LVOT D.
c
Position the calipers. (See "Working with 
calipers" on page 47.)
d
Save the calculation. (See "To save a 
calculation" on page 51.)
2 Measure from LVOT, and then measure from 
AV (Doppler):
• For Vmax, see "To measure peak velocity" 
on page 61. From the calculations menu, 
select AV, select sample site, and then 
select Vmax.
For VTI, see "To calculate Velocity Time 
Integral (VTI)" on page 61. From the 
calculations menu, select AV, select sample 
site, and then select VTI.
To calculate Qp/Qs
The Qp/Qs calculation requires two 
measurements in 2D and two measurements in 
Doppler. After the measurements are saved, the 
result appears in the patient report.
1 On a frozen 2D image, press the 
2 Do the following to measure from LVOT D 
and again to measure from RVOT D:
a From the calculations menu, locate Qp/Qs 
b
c
3 On a frozen Doppler spectral trace, press the 
CALCS
4 Do the following to measure from LVOT VTI 
and again to measure from RVOT VTI:
a From the calculations menu, select Qp/Qs 
b Press the 
c Using the touchpad, trace the waveform.
 key. 
CALCS
d Press the 
e
For information on the automatic trace tool, see 
"To trace automatically (Doppler)" on page 50.
To calculate Stroke Volume (SV) or Stroke
Index (SI)
The SV and SI calculations require a 
measurement in 2D and a measurement in 
Doppler. SI also requires Body Surface Area 
(BSA). After the measurements are saved, the 
result appears in the patient report.
1 (SI Only) Fill in the Height and Weight fields 
on the patient information form. The BSA is 
calculated automatically. (See "To create a 
new patient information form" on page 40.)
2 Measure from LVOT (2D):
 key.
a On a frozen 2D image, press the 
CALCS
b From the calculations menu, select LVOT D.
Chapter 4: Measurements and Calculations
and then select LVOT D or RVOT D.
Position the calipers. (See "Working with 
calipers" on page 47.)
Save the calculation. (See "To save a 
calculation" on page 51.)
 key.
and then select LVOT VTI or RVOT VTI.
 key to start the trace.
SELECT
To make a correction, select Undo 
on‐screen, backtrack with the touchpad, or 
press the 
 key.
BACKSPACE
 key to complete the trace.
SET
Save the calculation. (See "To save a 
calculation" on page 51.)
 key. 
CALCS
63

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents