To measure LVd and LVs
1 On a frozen 2D image or M Mode trace, press 
the 
 key. 
CALCS
2 From the calculations menu, select the 
measurement name.
3 Position the active (green) caliper at the 
starting point. (See "Working with calipers" 
on page 47.)
4 Press the 
 key, and position the second 
SELECT
caliper.
5 Press the 
 key.
SELECT
Another caliper appears, and the calculations 
menu highlights the next measurement name.
6 Position the caliper, and press the 
Repeat for each measurement name in the 
calculation group.
Each time you press the 
caliper appears, and the calculations menu 
highlights the next measurement name.
Save the calculation. (See "To save a 
7
calculation" on page 51.)
To measure Ao, LA, AAo, or LVOT D
1 On a frozen 2D image or M Mode trace, press 
the 
 key. 
CALCS
2 From the calculations menu, select the 
measurement name. 
3
Position the calipers. (See "Working with 
calipers" on page 47.)
4
Save the calculation. (See "To save a 
calculation" on page 51.)
To calculate LV Volume (Simpson's Rule)
1 On a frozen 2D image, press the 
2 Do the following for each measurement:
a From the calculations menu, select the 
desired view and phase.
b Position the caliper at the mitral annulus, 
and press the 
SELECT
60
Exam-based calculations
 key. 
SELECT
 key, another 
SELECT
 key.
CALCS
 key to start the trace.
c Using the touchpad, trace the left 
ventricular (LV) cavity.
To make a correction, select Undo on‐screen 
or press the 
BACKSPACE
d Complete the trace, and press the 
Save the calculation. (See "To save a 
e
calculation" on page 51.)
To calculate MV or AV area
1 On a frozen 2D image, press the 
2 In the calculations menu, locate Area, and 
then select MV or AV.
3 Position the caliper where you want to begin 
the trace, and press the 
4 Using the touchpad, trace the desired area. 
To make a correction, select Undo on‐screen or 
press the 
BACKSPACE
5 Complete the trace, and press the 
Save the calculation. (See "To save a 
6
calculation" on page 51.)
To calculate LV Mass
1 On a frozen 2D image, press the 
2 In the calculations menu, locate LV Mass.
3 Do the following for EPI and then for Endo:
a Select the measurement name from the 
calculations menu.
b Position the caliper where you want to 
begin the trace, and press the 
c Using the touchpad, trace the desired area.
To make a correction, select Undo 
on‐screen or press the 
d Complete the trace, and press the 
Save the calculation. (See "To save a 
e
calculation" on page 51.).
4 Select Apical from the calculations menu.