Chapter 4: Measurements and Calculations

You can measure for quick reference, or you can 
measure within a calculation. You can perform 
general calculations as well as calculations 
specific to an exam type.
Measurements are performed on frozen images. 
For references used, see Chapter 7, "References."

Measurements

You can perform basic measurements in any 
imaging mode and can save the image with the 
measurements displayed. (See "To save an 
image" on page 42.) Except for the M Mode HR 
measurement, the results do not automatically 
save to a calculation and the patient report. If you 
prefer, you can first begin a calculation and then 
measure. See "Performing and saving 
measurements in calculations" on page 51.
Some options may not apply to your system. 
Options available depend on your configuration, 
transducer, and exam type.
To save a measurement to a calculation and
patient report
1 With the measurement active (green), press 
the 
 key. 
CALCS
2 From the calculations menu, select a 
measurement name.
Only measurement names available for the 
imaging mode and exam type are selectable.
Save the calculation. (See "To save a 
3
calculation" on page 51.)
To start a calculation before measuring, see 
"Performing and saving measurements in 
calculations" on page 51.

Working with calipers

When measuring, you work with calipers, often 
in pairs. Results based on the calipers' position 
appear at the bottom of the screen. The results 
update as you reposition the calipers by using the 
touchpad. In trace measurements, the results 
appear after you complete the trace.
Outside a calculation, you can add calipers by 
pressing the 
 key. You can have multiple 
CALIPER
sets of calipers and can switch from one set to 
another, repositioning them as needed. Each set 
shows the measurement result. The active 
calipers and measurement result are highlighted 
green. A measurement is complete when you 
finish moving its calipers.
Within a calculation, calipers appear when you 
select from the calculations menu. (See "To select 
from the calculations menu" on page 51.)
For an accurate measurement, accurate 
placement of calipers is essential.
To switch the active calipers
Do one of the following:
• To switch the active caliper within a set, 
press the 
SELECT
• To switch the active set when measuring 
outside a calculation, select Switch 
on‐screen.
To delete or edit a measurement
With the measurement active (highlighted), 
do one of the following:
• To delete, select Delete on‐screen.
• To edit, use the touchpad to move the 
calipers.
Note: Trace measurements cannot be edited once set. 