When Either Normal Observation Mode Or Special Light Observation Mode Is Not Available; When Endoscopic Images Do Not Appear On The Monitor Or A Frozen Image Cannot Be Restored - FujiFilm 700 Series Operation Manual

Lower gastrointestinal endoscopes
Table of Contents

Advertisement

(8) Operate the up/down and left/right angulation knobs to straighten the bending section before
withdrawing the endoscope.
(9) Slowly withdraw the endoscope.
6.2.2
When Either Normal Observation Mode or Special Light Observation Mode
is Not Available
(1) Turn off all related equipment except the processor, light source, monitor, and suction pump.
(2) Operate the processor to switch to available observation mode.
(3) When using the optical zoom function, return the focusing position to "Normal."
(4) When using an endotherapy device, slowly withdraw the endotherapy device from the
endoscope.
(5) When using the flexibility adjustment mechanism, set the insertion tube to the most flexible
condition.
(6) Aspirate accumulated air (or CO
(7) Turn the up/down and left/right angulation locks in the direction of F (Free) until they stop to
unlock the up/down and left/right angulation knobs.
(8) Operate the up/down and left/right angulation knobs to straighten the bending section.
(9) Slowly withdraw the endoscope.
6.2.3
When Endoscopic Images Do Not Appear on the Monitor or a Frozen Image
Cannot be Restored
(1) Turn off the processor, light source, monitor, suction pump and other related equipment.
(2) When using an endotherapy device, slowly withdraw the endotherapy device from the
endoscope.
(3) When using the flexibility adjustment mechanism, set the insertion tube to the most flexible
condition.
(4) Turn the up/down and left/right angulation locks in the direction of F (Free) until they stop to
unlock the up/down and left/right angulation knobs.
(5) Operate the up/down and left/right angulation knobs to straighten the bending section and
release one's hand from the up/down and left/right angulation knobs.
(6) Slowly withdraw the endoscope.
897N201625
gas) by pressing the suction valve.
2
6-9