Defibrillator equipment must be maintained to be ready for immediate use. The defibrillator
should be tested daily. The R Series defibrillator gives you two readiness testing options:
Automatic
Manual
Note:
Maintenance Frequency
Visual Inspection
Code Readiness Test
(Automatic)
Manual Defibrillator
Testing
When the R Series device ships from ZOLL, the Code Readiness indicator may show a red
"X." A manual readiness test should be performed (in addition to other site-specific tests such
as HiPot and Leakage) prior to putting the device into service. Follow the procedure on page
12-3, and verify that the Code Readiness indicator displays a green "."
9650-0912-01 Rev. N
Both automatic and manual test results are automatically recorded to internal memory.
Once per day, inspect Code Readiness indicator.
Once per day, unless configured Off, in which case, perform manual
defibrillator testing daily.
Once per week; daily if Code Readiness Test is configured Off.
ZOLL R Series Operator's Guide
Chapter 12
Maintenance
12–1