Error Message
SYSTEM FAULT 36
SYSTEM FAULT 37
SYSTEM FAULT 38
TEST FAILED
TRANSFERRING DATA
USE PADDLE DISCHG
USE PADS TO PACE
USER SETUP REQ.
VF ALARMS OFF
ZERO CO2 ADAPTER
ZERO DONE
ZEROING CO2 ADAPTER
9650-0903-01 Rev. L
Explanation
Filtered sum of all the supply voltages is out of range.
1/2 scale reference voltage is out of range.
Failure to shutdown after "shutdown order" is written to the
RTC.
30J defib self-test failed.
Transferring data message.
Defibrillator is charged and front-panel discharge button
pressed when either external paddles or internal spoons
with discharge buttons are connected.
MFE accessory other than Pads detected in Pace mode.
Both copies of configuration data are bad, or software with
a configuration rev older than the current one was loaded.
Alarms enabled in Pace Mode or when Continuous
Analysis active or started in manual mode but the current
lead is not Pads, Lead I, Lead II, or Lead III. Also displayed
if the Alarm button is pressed but the Heart Rate alarm is
disabled.
Negative CO
detected. May be caused by a sensor that
2
was zeroed with CO
in the airway, or by an optical
2
blockage of the airway adapter.
The sensor/adapter zero is finished.
Adapter zeroing is in progress.
User
Technical Action
Advisory
Replace analog board or digital board.
Replace analog board or PD engine.
Replace digital board.
Replace universal cable, paddles or high voltage
module, capacitor, and/or system board.
Reconfigure unit.
Check the airway adapter and clean if necessary.
Perform a mainstream airway adapter zero as
described in "Zeroing the Mainstream
CAPNOSTAT 5 CO
2
No action required.
Wait for the adapter zeroing to finish.
R Series Service Manual
Sensor/Airway Adapter."
2-13