Wifi - ZOLL R Series Service Manual

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

R Series Service Manual
P/D Settings
The software attempts to establish a P/D setting that reflects an estimate of the battery ESR. This is to avoid shutdown due to loading of the DC bus. Tables
have been prepared on the basis of nominal R Series operating conditions without the printer operating. We will rely on the defib module throttling to
handle additional load conditions (such as printer).

WiFi

The R Series offers the option for WiFi data transmission of the full disclosure file (ECG waveforms, SpO
values, etc.), device check file (self test
2
results), and activity log file. This is accomplished with an IEEE 802.11abgn Wi-Fi compact flash card that is inserted into the existing compact flash I/O
port. Full disclosure transmissions can be sent (when properly configured) to a database management system called CodeNet Central that resides on a
desktop or laptop computer. Device check and activity log transmissions will be sent to a ZOLL Data Systems application called Defib Dash Board.
CodeNet Central is developed by ZOLL Data Systems.
5-6
9650-0903-01 Rev. L

Hide quick links:

Advertisement

Table of Contents
loading