R Series ALS/ Plus In-service Script

Device
-Color Coded Operating System: Start with the dial. Three different colored
coded areas.
RED
for Defib,
-Setup on Code Cart
-An Adult OneStep CPR Electrode should ALWAYS be connected to the therapy
cable.
-This allows "Code-Readiness" and allows the device to pass its Automated
Daily Test.
-Power LED lights
1. Battery Indicator: GREEN=Charged, AMBER=Charging
FLASHING=Problem (Battery Check)
2. A/C Power Indicator GREEN=Device is plugged in
3. No lights=Not plugged
-Code Readiness Indicator:
Ready
-Self Test- Every
device
is ready for use, or
-Failed Test- The device will display "Readiness Test Failed" on the screen and will
-These are usually easily fixable problems (No pads plugged in, no battery, no A/C power, etc)
-If the device reads "DEFIB FAILURE", call BioMed.
-Battery- Indicator lights show 30 min. incriminates for total of 4 hours of runtime.
- If orange light appears on "?" section of battery, call BioMed
-Recorder Tray
Press button to open door, insert the paper with the black arrow facing towards the back of the
device. Pull a strip or two forward and let paper drop. Then close the door on top of the strip.
-Cables/Caps
1. Electrotherapy Cable- (Is bolted down)
2. SpO2 (Blue/If Applicable)
3. 3 Lead EKG/One-Step Cable
4. EtCO2 (Yellow/If Applicable)
5. NIBP- Non-Invasive Blood Pressure (Black/If Applicable)
6. Cable Compartment for all cables attached.
One-Step Cable: Comes out the side and is 8 feet in length.
GREEN
for Pacing,
in
GREEN Check mark= Code Ready
24 hours, the device will do a 100 point test and
if it failed a part of the test.
highlight what needs to be fixed.
GREY
for monitoring.
RED "X"= Not Code
will display a if the

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the R Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ZOLL R Series

  • Page 1 R Series ALS/ Plus In-service Script Device -Color Coded Operating System: Start with the dial. Three different colored coded areas. for Defib, GREEN for Pacing, GREY for monitoring. -Setup on Code Cart -An Adult OneStep CPR Electrode should ALWAYS be connected to the therapy cable.
  • Page 2 Using the Device- Everything will be pre connected and ready for use. This means it is ready for a CODE and TESTING. Connecting and Disconnecting Pads Locking lever holds pads and cable into place. Press on the far end of the lever to release. Using the Pads It is important to prep the skin as best as you possibly can.
  • Page 3 Converting To Manual Mode (R Series Plus Model ONLY) Depending on how your facility configures the device, you can convert the R Series Plus in to Man- ual Mode by one of the two ways: Press the Manual Mode Key (bottom left corner) Press the Manual Mode key, and then the Confirm Key (bottom right corner) (Simulator should be turned to “VT HI”)
  • Page 4 Pediatric Feedback: Will only display Rate and Depth- you will not hear the “Push Harder” voice prompt. —————————————-BLS USERS ARE NOW EXCUSED FROM TRAINING——————————— See-Thru CPR Technology (Simulator is Changed to VF Artifact) -Two Wave Forms 1) Raw EKG-Same waveform with artifact 2) Filtered EKG- Internal algorithm that filters out any artifact allowing you to see the underlying rhythm.
  • Page 5 Monitor Mode -Lead Key: Can change the lead by pressing lead- Circulates through Lead I-III and PADS -Size Key: The gain of the trace. Makes the waveform bigger or smaller. -Alarm Suspend: Alarms a default to be “Off”. (Unless otherwise Configured) 1.
  • Page 6 Test Log The R Series can hold the last 999 Tests performed (manual or automatic) and can be reviewed/printed out for verification at anytime. Test Log will never become “full” as once the memory reaches 999 and a new test is entered, the oldest test will be deleted to make room.

Table of Contents