Applying Therapy Proper Electrode or Paddle Placement OneStep Complete Resuscitation Electrode OneStep CPR Anterior/Anterior Anterior/Posterior Placement Electrode Placement FRONT BACK FRONT BACK Atrial Arrhythmias Ventricular Arrhythmias Cardioversion Electrode Cardioversion Electrode Anterior/Posterior Placement Anterior/Posterior Placement FRONT BACK FRONT BACK Paddle Placement 1.
Adult Defibrillation R Series BLS/Plus Model Defibrillation (AED) Defibrillation (Advisory) 1. Apply electrodes. 1. Apply electrodes. 2. Turn selector switch to ON. 2. Turn selector switch to ON. 3. Press Manual then Confirm softkeys. 3. System will begin to analyze or instruct PERFORM CPR, depending 4.
8 years of age (55lbs/25kg). For children over 8 years of age, use adult electrodes. • When the OneStep Pediatric electrode is connected, the R Series will automatically set the energy delivered to 50-, 70-, and 85-joule increments. These levels are appropriate for children from 55lbs/25 kg or less.
Release Bar (adult pads only) (adult pads only) CPR Dashboard with Real CPR Help Rate and Depth Indicators (adult and pediatric pads) See-Thru CPR See-Thru CPR is a proprietary algorithm that filters the CPR artifact from the ECG to allow users to see a close approximation of a patient’s organized underlying rhythm Unfiltered ECG signal during CPR during CPR compressions, reducing interruptions in CPR.
4. Turn the pacer OUTPUT knob clockwise slowly until capture. Dial settings for 5. Slowly decrease output by turning OUTPUT knob counter- R Series Pacing clockwise until capture disappears, then increase by 10%. 6. Check pulse to verify mechanical capture.
6. Apply electrodes to patient’s skin. ECG Display 1. Turn selector switch to MONITOR on the R Series ALS unit. If using BLS or Plus device, turn ON, then press Manual then Confirm softkeys to activate ALS mode. 2. Press LEAD button to change leads.
Page 10
SpO 2 port on the rear of the unit labeled SpO 2 . 2. Attach sensor to patient’s finger. 3. Turn selector switch to grey area labeled MONITOR on the R Series ALS. If using BLS or Plus device, turn ON, then press Manual then Confirm softkeys to use in ALS mode.
3. Press the Print Log softkey Transferring Full Disclosure Files to a Compact Flash Card or by WiFi: 1. If using an R Series ALS, turn to MONITOR mode 2. Press the Report Data softkey. 3. Press the Transfer Mode softkey.