Delivered Energy At Every Defibrillator Setting Into A Range Of Loads - ZOLL R Series Service Manual

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

R Series Service Manual
Selected
Energy
1 J
2 J
3 J
4 J
5 J
6 J
7 J
8 J
9 J
10 J
15 J
20 J
30 J
50 J
70 J
75 J
85 J
100 J
120 J
150 J
200 J
9650-0903-01 Rev. L
Delivered Energy at Every Defibrillator
Setting into a Range of Loads
25
50
75
1 J
1 J
1 J
1 J
2 J
2 J
2 J
3 J
3 J
3J
4 J
4 J
3 J
5 J
6 J
4 J
6 J
7 J
5 J
7 J
8 J
5 J
8 J
9 J
6 J
9 J
10 J
7 J
10 J
12 J
10 J
16 J
17 J
14 J
21 J
23 J
21 J
32 J
35 J
35 J
54 J
59 J
49 J
76 J
83 J
53 J
81 J
89 J
60 J
92 J
101 J
71 J
109 J
119 J
85 J
131 J
143 J
107 J
164 J
180 J
142 J
230 J
249 J
* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is greater.
Load
100
125
150
175
1 J
1 J
1 J
2 J
2 J
2 J
3 J
3 J
3 J
5 J
5 J
5 J
6 J
6 J
6 J
7 J
7 J
7 J
8 J
8 J
8 J
9 J
10 J
9 J
11 J
11 J
11 J
10 J
12 J
12 J
12 J
12 J
18 J
18 J
18 J
17 J
24 J
24 J
24 J
23 J
36 J
37 J
36 J
35 J
61 J
62 J
61 J
59 J
85 J
87 J
86 J
83 J
91 J
93 J
92 J
89 J
104 J
106 J
104 J
101 J
122 J
125 J
123 J
119 J
147 J
150 J
147 J
143 J
183 J
188 J
184 J
179 J
253 J
269 J
261 J
260 J
Accuracy*
1 J
±15%
2 J
±15%
3 J
±15%
4 J
±15%
6 J
±15%
7 J
±15%
8 J
±15%
9 J
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
A-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents