ZOLL R Series Service Manual page 62

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

R Series Service Manual
Error Message
NIBP FAULT 263
NIBP FAULT 264
NIBP FAULT 265
NIBP FAULT 266
NIBP FAULT 267
NIBP FAULT 268
NIBP MEAS ABORTED
NIBP NOT READY
NIBP OUT OF RANGE
NIBP SIGNAL WEAK
NO DATA TO TRANSFER
2-8
Explanation
Fault 90 received from NIBP module.
Fault 91 received from NIBP module.
Fault 97 received from NIBP module.
Fault 98 received from NIBP module.
Fault 99 received from NIBP module.
Device no response after power up.
• Cuff inflation pressure is set too high for attached cuff.
• Inflation is too fast.
• R Series is unable to find systolic value for 180
seconds.
• Defibrillator is charged or charging.
• User initiated abort.
• The defibrillator is charged or charging in progress.
• NIBP module is performing power-up self-test.
The data from the NIBP module is out of range.
There is a weak or no oscillometric signal.
Summary Report / Device Check / Activity Log data
transfer error.
User
Technical Action
Advisory
Replace NIBP module.
Replace NIBP module.
Replace NIBP module.
Replace NIBP module.
Replace NIBP module.
Replace NIBP module.
• Verify that you are using proper size cuff.
• Check for cuff and hose blockage.\
• Confirm that the unit was not charging.
• If the problem persists, contact ZOLL
Technical Support.
• Wait until the unit discharges before taking the
next measurement.
• Wait for more than 10 seconds after power-up
before taking blood pressure measurements.
• Check cuff fit and positioning. Switch cuff to
other arm.
• Measure patient's blood pressure with other
equipment.
• If the problem persists, contact ZOLL
Technical Support.
• Check cuff fit and positioning.
• Check for kinked hose.
• Increase cuff inflation pressure if clinically
appropriate.
9650-0903-01 Rev. L