Heart Rate Display Test - ZOLL R Series Service Manual

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

7.0 Heart Rate Display Test

Tools Needed:
Impulse 4000
ECG Cable (3 or 5 leads).
Test Setup:
1. Turn the selector switch to MONITOR (for AED units turn to ON.) Press LEAD button until "I" displays.
Connect the ECG leads to the Fluke Biomedical 4000 or equivalent.
2.
Connect the ECG cable to the unit.
3.
Do this...
7.1
Set the Impulse 4000 to 120BPM.
Record your results on the Maintenance Test Checklist.
9650-0903-01 Rev. L
Observe this...
The Heart Rate displays as 120 +/- 2 bpm.
R Series Service Manual
Pass/Fail/NA
o o o
17
1-