To disable detection of pacemaker spikes:
1. Press the Param softkey.
Press ECG.
2.
Press Disable Pacer Detect.
3.
The Pacer Off icon (
implanted pacemaker pulses by the heart rate detector is disabled. Do not disable the implanted
pacemaker detector when monitoring patients known or suspected to have an implanted pacer.
Disabling pacemaker pulse detection under these conditions can result in erroneously counting
pacemaker pulses as R-waves leading to the display of inaccurate heart rates.
To re-enable detection of pacemaker spikes:
1. Press the Param softkey.
2.
Press ECG.
Press Enable Pacer Detect.
3.