R Series Service Manual
19.0 EtCO
Monitor Test (for EtCO
2
Tools Needed: CAPNOSTAT 5 Mainstream cable with airway adapter.
Test Setup:
1. Install the battery.
Do this...
19.1
Connect the CAPNOSTAT 5 CO2 Mainstream
cable with airway adapter attached to the yellow
connector at the back of the R Series.
19.2
Set the front panel switch to MONITOR or ON.
For AED units, enter Manual Mode.
19.3
When the WARM UP message disappears,
press the Param softkey, then select EtCO2.
19.4
Press the ZERO softkey, then wait for the
ZERO DONE message.
19.5
Press the Return softkey.
19.6
Press the following softkeys to display the CO
waveform, Options, Traces, Trace 2, then
EtCO2,
19.7
Breathe normally into the airway adapter.
Record your results on the Maintenance Test Checklist.
1-34
Option)
2
Observe this...
NOTE: Make sure the airway adapter is installed in the CO
WARM UP message appears on the display.
NOTE: Warming up may take up to 3 minutes.
The ZERO DONE message appears.
A flat baseline CO
waveform appears.
2
2
A capnogram waveform appears.
Pass/Fail
cable.
2
o o
o o
o o
o o
9650-0903-01 Rev. L