Error Message
ENERGY INCREMENTED
ERASING REPORT
FULL DISCLS STOPPED
IF NO PULSE
INSERT CARD
INT. DUMP OVERLOAD
INVALID GAS TEMP:
RETRY
LOW BATTERY
NIBP ARTIFACT
NIBP COMM ERR 259
NIBP COMM ERR 260
NIBP COMM ERR 261
NIBP COMM ERR 262
9650-0903-01 Rev. L
Explanation
Defib energy has been automatically incremented to the
next configured level after shock 1 or 2 has been delivered
and the unit is configured for Basic Energy Auto
Escalation.
The unit is erasing the selected report data.
Full disclosure data exceeds the storage capacity: 4 hours
or 32Mb.
Message displayed upon entered CPR period specified by
AHA protocol.
No card detected during manual or semi-automatic modes.
More than 15 internal discharges in 5 minutes.
The operator has attempted to set a gas temperature
outside the Capnostat's operating range.
Low battery.
The unit is unable to detect systolic, diastolic or mean
blood pressure due to excessive motion or vibration.
Framing, parity, or fifo error.
Received invalid packets.
Device not taking action request.
No reply from device.
User
Technical Action
Advisory
Re-insert CF Card or replace communication
module.
Unit needs to cool down, wait approximately
5 minutes.
Recalibrate EtCO
, verify range.
2
• Replace battery pack with a fully charged
battery pack.
• Plug unit into AC mains.
• Take a single blood pressure measurement.
• Keep patient as still as possible.
• Insulate patient, cuff and hose from vibrations
as much as possible.
Replace parameter power supply or NIBP module.
Replace parameter power supply or NIBP module.
Replace parameter power supply or NIBP module.
Replace parameter power supply or NIBP module.