C
1 M
HAPTER
AINTENANCE
Preparations for Testing
•
Gather the required equipment listed in the next section.
•
Perform all the steps within each test procedure in the order specified.
If testing results indicate that the battery pack must be replaced, you will need a fully charged
spare battery pack.
Test Equipment
For some tests, you need one or more of these items:
•
ZOLL defibrillation analyzer (universal) adapter cable (8000-0804-01)
•
Defibrillator Analyzer (e.g. QED 6) or Fluke Impulse 4000 Defibrillator/Pacer Performance
Analyzer (Fluke Biomedical/ DNI Nevada) with 1.06 software or higher
•
Battery Current Test Fixture (P/N 9100-0690-TF),
•
Power Supply 12V 10A,
•
AED Plus
•
DMM (e.g. Agilent 34401A Digital Multimeter) or equivalent.
•
AED Pro Disposable Battery (P/N 1008-002003-01)
•
SurePower
•
PC with ZOLL RescueNet
7771-000012-01),
•
ZOLL USB IrDA port (P/N 9355-0807).
•
PC with AED Plus/PRO DCI Panel software (P/N 9103-0815-SW),
•
AED Pro Resistance, Decade Box (P/N 9100-000957-TF) or equivalent (e.g.RS-200
Resistance Substituter).
•
ZOLL 3-lead ECG cable (8000-0838 or 9500-001802)
•
USB Flash Drive (P/N 8000-000474),
For ZOLL part numbers, refer to the accessories list in the AED Pro Operator's Guide.
1–2
T
ESTS
®
, Quick-Test, Test Fixture (P/N 9100-0567-TF)
®
Rechargeable Li-Ion Battery (P?N 1009-001535-03)
®
Code Review and ZAS (ZOLL Administration Software, P/N
www.zoll.com
9650-0309-01 Rev. E