Select On - ZOLL R Series Service Manual

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

C
4 M
HAPTER
ANUAL

1 Select ON

ON
OFF
The unit beeps 4 times to indicate that it has passed the power on self test, and the green AED
label and the ON mode selector light up on the front panel.
Press the Manual Mode softkey on the front panel of the unit to enter the Manual mode of
operation.
The message CONFIRM MANUAL MODE and the Confirm softkey will be displayed. Press
the Confirm softkey to enter the Manual mode of operation. If the Confirm softkey is not
pressed within 10 seconds, the unit will revert back to AED operation.
Note:
The control buttons for Manual mode light up on the front panel and the unit automatically
defaults to 120 joules or the preconfigured first shock energy selection.
PADS are selected as the ECG source when the instrument is turned to Manual mode and
paddles are not connected to the OneStep cable. You may select any of the other ECG leads by
pressing the front panel LEAD button.
Energy Select
Look at the display, and verify the selected energy is appropriate. The default energy selections
for adult patients are:
Shock 1 - 120 joules
Shock 2 - 150 joules
Shock 3 - 200 joules
When used with OneStep Pediatric electrodes, the default energy selections are:
Shock 1 - 50 joules
Shock 2 - 70 joules
Shock 3 - 85 joules
WARNING!
When used with other pediatric defibrillation electrodes, defibrillator energies must be
set manually based on site-specific institutional protocols for pediatric defibrillation.
After the third shock, all subsequent shocks are delivered at the same energy as the third shock
in both Adult and Pediatric modes.
4–8
D
EFIBRILLATION
AED
If the unit has been configured not to confirm the transition from AED mode to
Manual mode, the unit will immediately enter Manual mode.
www.zoll.com
9650-0904-01 Rev. N