ZOLL R Series Service Manual page 526

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

A
A S
PPENDIX
PECIFICATIONS
Table A-2. Delivered Energy at Every Defibrillator Setting into a Range of Loads
Selected
Energy
25
150 J
107 J
200 J
142 J
The R Series Rectilinear Biphasic waveform employs the same first and second phase timing,
similar first and second phase currents/voltages, and essentially the same mechanisms for
controlling defibrillation waveshape as the ZOLL M Series
R Series defibrillation waveforms are considered substantially equivalent.
Figures A-1 through A-21 show the Rectilinear Biphasic waveforms that are produced when
the R Series defibrillator is discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at
each energy setting.
The vertical axis shows the current in amperes (A); the horizontal axis shows the duration of
time in milliseconds (ms).
40
30
20
10
0
0
-10
-20
-30
-40
Figure A-1. Rectilinear Biphasic Waveform at 200 Joules
A–14
Load
50
75
100
164 J
180 J
183 J
230 J
249 J
253 J
2
4
www.zoll.com
125
150
188 J
184 J
269 J
261 J
®
defibrillator. The M Series and
6
8
ms
175
Accuracy
179 J
±15%
260 J
±15%
25 Ohm
50 Ohm
75 Ohm
100 Ohm
125 Ohm
10
12
150 Ohm
175 Ohm
200 Ohm
9650-0912-01 Rev. N