Code Readiness Log

Each automatic and manual defibrillation test result is stored in internal non-volatile memory,
the Code Log. A total of 1000 Code Readiness test records can be stored in internal memory.
When the Code Readiness Log is full, the oldest records are erased on a first-in-first-out (FIFO)
basis.
The Code Readiness Log can be transferred to an external computing device (see "Transferring
Files to an External Device" on page 12-1).
If configured, the R Series prints a Code Readiness Test Report following the completion of
each automatic defibrillator test.
To print the Code Readiness Log:
1. Turn the Mode Selector to ON.
2.
Press the Manual Mode softkey.
Note:
Press the Report Data softkey.
3.
4.
Press the Test Log softkey.
A menu appears with the print options.
9650-0904-01 Rev. N
If the CONFIRM MANUAL MODE message displays, press the Confirm softkey to
enter the Manual mode of operation.
ZOLL R Series Operator's Guide
Manual Defibrillator Testing
13–7