The following figure shows a patient in VF, which, during compressions, is slightly more
difficult to discern. When viewing this ECG, it is possible to view the underlying rhythm as the
filter is able to reject all of the CPR artifact.
CoarseVF
Raw ECG
Filtered ECG
CPR
0:00
CoarseVF
Raw ECG
| Filter ON
Filtered ECG
CPR
0:12
CoarseVF
Raw ECG
| Filter OFF
Filtered ECG
CPR
0:24
8–4
CPR (R S
P
ERIES
LUS
www.zoll.com
O
)
NLY
Raw ECG
Filtered ECG
CPR
0:12
Raw ECG
Filtered ECG
CPR
0:24
Raw ECG
Filtered ECG
CPR
0:36
12.5 mm/sec, 5 mm/mV
9650-0904-01 Rev. N