Preparations for Testing

Gather the required equipment listed in the next section.
Perform all the steps within each test procedure in the order specified.
If testing results indicate that the batteries must be replaced, you will need a set of 10 fully
charged Type 123A batteries.

Test Equipment

For some tests, you need one or more of these items:
ZOLL universal adapter cable
Fluke
DNI Nevada) with 1.06 software or higher
The following items are optional:
ZOLL AED Plus Simulator —
When connected to an AED Plus unit, the simulator can send it various simulated cardiac
rhythms (ventricular fibrillation, normal sinus rhythm, or asystole) and optional simulated
chest compression readings (rate and depth), so that you can test the unit's response.
ZOLL Administration Software (ZAS) —
When installed on a personal computer or personal digital assistant that has an IrDA
(wireless infrared) interface, the ZAS application can communicate with an AED Plus unit
in nonrescue mode to retrieve its usage history, hardware and software configuration, and
error log. You can also use ZAS to view or modify the unit's configurable settings and to
retrieve event data.
For ZOLL part numbers, refer to the accessories list in the AED Plus Administrator's Guide.
1–2
T
ESTS
®
Impulse 4000 Defibrillator/Pacer Performance Analyzer (Fluke Biomedical/
www.zoll.com
®
9650-0308-01 Rev. A