Error Message
CO2 UNIT ERROR
CO2 WARM UP
CPR FAULT 8
DATA TRANSFERRED
DEFIB DISABLED
DEFIB FAULT 76
DEFIB FAULT 77
DEFIB FAULT 78
DEFIB FAULT 79
DEFIB FAULT 80
DEFIB FAULT 94
DEFIB FAULT 95
DEFIB FAULT 96
DEFIB MAINT. REQ.
9650-0903-01 Rev. L
Explanation
The EtCO
sensor or module has detected a hardware
2
error.
The mainstream sensor is warming up. This may take up
to 5 minutes.
ECG processor not receiving CPR data
Transfer done message
User prompt issued simultaneously with other faults if defib
is disabled.
PD Defib failure during POST.
PD Defib failure while running.
PD Defib functional safety error while running.
PD module Defib/Pace failure during POST.
Undefined error received from the PD module.
PD module reset 3 times without being requested to.
PD module not communicating.
PD module reported an error on the discharge.
More than 5000 discharges of 200J have occurred.
Maintenance is required.
User
Technical Action
Advisory
Check that the sensor is properly plugged in.
Re-insert the sensor. Turn R Series unit off, then
on again to reset. perform a mainstream airway
adapter or module zero, as described in "Zeroing
the Mainstream CAPNOSTAT 5 CO
Airway Adapter".
If the problem persists, contact ZOLL Technical
Support.
Wait for sensor or module to warm up.
If the message persists more than 5 minutes,
replace the sensor.
Replace analog board.
Possible configuration problem. Replace high
voltage module. Call ZOLL Technical Support.
Replace PD engine.
Replace PD engine.
Replace PD engine.
Replace PD engine.
Replace PD engine.
Replace PD engine.
Replace PD engine.
Replace PD engine.
Replace PD engine.
R Series Service Manual