Error Message
DEFIB NOT CHARGED
DEFIB OVERUSE
DEFIB PAD SHORT
DISK FORMAT REQ.
ECG DISABLED
ECG FAULT 200
ECG FAULT 3
ECG FAULT 4
ECG FAULT 5
ECG FAULT 6
ECG FAULT 7
ECG LEAD OFF
ECG RA LEAD OFF
ECG TOO LARGE
ECG V LEAD OFF
ENABLE ETCO2
2-6
Explanation
Discharge button is pressed in a Defib mode but the unit is
not charged.
More than 50 shocks were delivered in less than 20
minutes.
Measured impedance between high voltage leads of MFC.
Report error if any problem with DOC file access occurred.
Persistent Critical Hardware Failure on ECG module.
No ECG applicable available.
ECG processor reset failure.
Excessive number of missed samples from ECG
processor.
ECG processor power up failure.
ECG processor Hardware failure.
ECG processor is not responding.
1 or more ECG leads are not connected when non-MFE
leads are selected as input.
ECG lead RA is disconnected
Number of intervals with large ECG exceeds threshold.
Issued immediately before RETRY ANALYSIS.
ECG lead V is disconnected
User attempts to zero EtCO
sensor while the EtCO
2
feature is disabled
User
Technical Action
Advisory
Unit needs to cool down, wait approximately
5 minutes.
Ensure pads are coupled to patient. Check/
replace pads or universal cable. Replace system
board.
Replace digital board.
Replace analog board.
Reload MCU Software and ECG App. Software.
Replace analog board.
Replace analog board.
Replace analog board.
Replace analog board.
Replace analog board.
Using the softkeys, select Enable EtCO2.