ZOLL R Series Service Manual page 31

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

Do this...
10.16
Remove a lead wire from the Impulse 4000.
10.17
Reattach ECG Lead wire to Impulse 4000 and
hold the ALARM SUSPEND button on unit for 4
seconds.
10.18
Press the ALARM SUSPEND button.
10.19
Set simulator to 160 BPM or higher.
10.20
Press the ALARM SUSPEND button in the unit.
10.21
Press and hold ALARM SUSPEND for
3 seconds to disable alarms.
Record your results on the Maintenance Test Checklist.
9650-0903-01 Rev. L
Observe this...
The bell symbol flashes and the heart symbol stops flashing. The ECG LEAD
OFF alarm tone sounds and the Heart Rate Value is highlighted. Recorder
prints a stripchart showing a low heart rate, if enabled.
The bell symbol has an "X" through it.
The heart symbol flashes with each QRS wave.
Alarm is enabled. Bell symbol (without "X") displays.
Heart Rate Value is highlighted, alarm tone sounds, the bell and the heart
symbol both flash.
Alarm is suspended for 90 seconds. The bell symbol has an "X" through it. The
heart symbol flashes with each QRS wave and the Heart Rate Value is
highlighted.
Verify the Heart Rate value is not highlighted, and the bell symbol displays with
an "X" through it and is not flashing.
R Series Service Manual
Pass Fail/NA
o o o
o o o
o o o
o o o
o o o
o o o
21