Operator's Checklist for R Series Product

Recommended checks and procedures to be performed
daily or weekly, depending on Automatic Test
configuration.
1. Condition
Unit clean, no spills, clear of objects on top, case intact
2. Hands-free Therapy electrodes
1 set preconnected / 1 spare
3. Paddles
Paddles clean, not pitted
Release from housing easily
4. Inspect cables for cracks, broken wires, connector
A ECG electrode cable, connector
B Defibrillator paddle cables
C OneStep cable, connector
D Other patient cables
5. Batteries/External power supply
A Fully charged battery in unit
B Fully charged spare battery available, or ready access to local AC mains
6. Disposable supplies
A Defib gel or gel patches
B Hands-free therapy electrodes in sealed packages — 2 sets
C ECG electrodes
D Recorder paper
E Alcohol wipes
F Razors
7. Operational checks
A Power On Sequence
Turn unit to ON, 4-beep tone heard, AED label and ON Mode Selector light up
Press Manual Mode softkey, press Confirm softkey, front panel lights up
ECG size X 1
"PADDLES" or 'PADS" as lead selected
B Paddles
Paddles in holder: Set defib energy level to 30 joules, press paddles firmly into the side
wells, and simultaneously press and hold both defib discharge buttons; "30J TEST OK"
message on Recorder.
C Defibrillator
OneStep cable connected to test connector,
level to 30 joules, press SHOCK button; "30J TEST OK" message on Recorder
D Pacer Operation (If installed)
OneStep cable not connected to Test Connector or OneStep electrodes
Turn to PACER, set pacer rate to 150 ppm, press RECORDER button
Pacer pulses occur ever 2 large divisions (10 small divisions)
Press 4:1 button, pulses occur every 8 large divisions
Set PACER OUTPUT to 0 mA, no CHECK PADS message
Set PACER OUTPUT to 16 mA, CHECK PADS message and alarm
Reconnect OneStep cable to test connector,
Alarm softkey; CHECK PADS message disappears and Pace alarm stops.
E Recorder
Press RECORDER button; Recorder runs. Press again; Recorder stops.
Inspect Recorder printing
8. Please check the appropriate box after each use of this checklist.
No action required
Minor problem(s) corrected
Disposable supplies replaced
Major problem(s) identified — UNIT OUT OF SERVICE
Date ______________________________________________
Location ____________________________________________
Unit Serial Number ___________________________________
or OneStep electrodes.
Set defib energy
or OneStep electrodes
. Press Clear Pace
Remarks
Signature
______________________