Recorder; Pacer - ZOLL R Series Service Manual

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

C
14 T
HAPTER
ROUBLESHOOTING

Recorder

Symptom
1. CHECK RECORDER message
2. Recorder makes stuttering sound
when activated.
3. Light or poor quality printing
4. Summary Report will not print
when Report/Print Chart softkeys
are activated.

Pacer

Symptom
1. CHECK PADS message
2. No stimulus marker (
present on the ECG trace.
3. No ventricular capture beat
appears after stimulus marker on
ECG display.
14–4
Recommended Action
• Recorder out of paper.
• Remove paper, check paper type, check recorder for
paper jam, and reload paper.
• Recorder door is open.
• Paper is loaded upside down.
• Check recorder for paper jam.
• Ensure correct paper is in use.
• Ensure paper is installed grid side against recorder print
head.
• Recorder print head requires cleaning.
• 15 seconds have not elapsed since one of the events
that trigger Summary Report to record have occurred.
Wait 15 seconds, then try again.
Recommended Action
• Ensure therapy electrodes are connected to the
OneStep cable.
• Ensure electrode gel is not dry. Replace therapy
electrodes if necessary.
• Ensure good electrode-to-patient contact.
• Check integrity of OneStep cable by plugging into test
connector. CHECK PADS should disappear.
• Ensure unit is in PACER mode.
) is
• Ensure PACER RATE (ppm) is set greater than patient
heart rate.
• Check patient's pulse.
• Increase output current.
• Ensure therapy electrodes are making good contact
with the patient.
• Select different ECG Lead configuration.
• Review therapy electrode placement.
www.zoll.com
9650-0904-01 Rev. N