ZOLL R Series Service Manual page 515

Defibrillator
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

Messages
Electrodes
Hands-free therapy
electrodes
Defibrillator
Waveform
Energy selection
(delivered to a 50
load)
Charge Time
Patient Impedance
Range
Energy display
Synchronized
mode
Charge controls
Paddles
Automatic
Defibrillator Test
9650-0912-01 Rev. N
ERASING REPORT, INSERT CARD, REPLACE BATTERY,
LOW BATTERY, PERFORM CPR, NOISY ECG, RETRY ANALYSIS,
CHECK PATIENT, ANALYSIS HALTED, PRESS ANALYZE,
NO SHOCK ADV., SHOCK ADVISED, PRESS CHARGE, SELECT PADS,
SELECT ECG LEADS, SELECT DEFIB MODE,
VF ALARMS OFF, REMOVE SYNC, CHECK PADS, ATTACH PADS,
POOR PAD CONTACT, DEFIB PAD SHORT, PADDLE FAULT,
ECG LEAD OFF, USE PADDLE DISCHG, CANNOT CHARGE, RELEASE
SHOCK, PRESS SHOCK, 30J TEST OK, TEST FAILED, PACER
DISABLED, DEFIB DISABLED, SET PACE MA,
CHECK RECORDER, ANAYLZING ECG, FULLY RELEASE
• OneStep electrodes
-
• Pro
padz
-
• Stat
padz
®
• Pediatric pedi-padz
Rectilinear Biphasic
• ADULT: 1, 2, 3, 4, 5, 6, 7 8, 9, 10, 15, 20, 30, 50, 75, 100, 120, 150,
200 joules (through front panel buttons or paddle buttons)
• PEDIATRIC: 1, 2, 3, 4, 5, 6, 7 8, 9, 10, 15, 20, 30, 50, 70, 85, 100, 120,
150, 200 joules (through OneStep pediatric electrodes only)
• Less than 7 seconds with a new, fully charged battery (first 15 charges
to 200 joules), with or without AC power at rated mains voltage.
• For the fifteenth discharge at maximum energy, the charge time is less
than 10 seconds, with or without AC power at rated mains voltage.
Depleted batteries result in a longer defibrillator charge time.
• Less than 15 seconds when operating without a battery, using AC
power alone at 90% of the rated mains voltage.
• Less than 25 seconds from the initial power on, with a new, fully
charged battery pack (depleted by up to fifteen 200 Joule discharges)
or when operating without a battery, using AC power alone at 90% or
100% of the rated mains voltage.
• Less than 30 seconds from initial power on and after rhythm analysis
(advisory mode) with a new, fully charged pack (depleted by up to
fifteen 200 J discharges) and using AC power at 90% of rated mains
voltage.
15 - 300 ohms
Screen shows selected and delivered energy.
Synchronizes defibrillator discharge to the patient's R wave. SYNC is
indicated on the display with R wave markers above the ECG waveform on
the screen and stripchart. When ECG is monitored by the device, meets
the DF-80:2003 requirement of 60ms maximum time delay between the
peak of the R wave and the delivery of energy.
CHARGE button on front panel and apex paddle.
Standard apex/sternum paddles. Adult plate slides off to expose smaller
plate for pediatric patients.
Verifies defibrillator charging and discharging without removing paddles
from storage wells or with OneStep cable connected to the Test Port or
OneStep electrodes.
ZOLL R Series Operator's Guide
Defibrillator Specifications
A–3

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for ZOLL R Series

Table of Contents