Do this...
17.5
Increase the output to 80mA
17.6
Increase the output to 120mA.
17.7
Increase the output to 140mA.
17.8
Decrease the output to 60mA.
Decrease the rate to 30 ppm.
17.9
Connect the ECG cable to the R Series and
Impulse 4000. Select the ECG at 60 BPM on
the Impulse 4000. Increase the pacer rate on
the unit to 58ppm.
17.10
Press the Async Pace softkey.
17.11
Turn off Impulse 4000. Set Pacer Rate to
100ppm. Press the RECORDER ON button.
17.12
Press and hold 4:1 button.
Record your results on the Maintenance Test Checklist.
9650-0903-01 Rev. L
Observe this...
Output on the Impulse 4000 is 80mA or +/- 5mA.
Output on the Impulse 4000 is 120mA or +/- 6mA.
Output on the Impulse 4000 is 140mA or +/- 7mA. Pulse width is 40mS +/-
2mS. Pacer rate on Impulse 4000 is 177-183 ppm.
Pacer rate on Impulse 4000 is 29-31 ppm.
ECG at 60 BPM is seen on the display and no stimulus markers.
ECG at 60 BPM seen on the display with the pace stimulus markers displayed.
Async pace message displays.
Observe the pace stimulus markers every 15mm +/-1mm.
Observe the pace stimulus markers every 60 mm+/- 1.5 mm.
R Series Service Manual
Pass/Fail
o o
o o
o o
o o
o o
o o
o o
o o
31
1-